Description
Offers 3 call levels with distinct tone and light signaling for routine, urgent, and cord-out trouble calls.
Features
- Routine call and bathroom call annunciated by slow or fast flashing LED and tone
- Auto-connect to next call-in priority (no need to select station)
- Fault tone sounds when wires are disconnected or shorted at patient station, or handset is off-hook for more than 90 seconds
Specifications
- Communication: Handset (VOX) or push-to-talk (simplex)
- Call-In Priority:
- Cord-out trouble call
- Urgent bathroom call and NHR-3TS common area bathroom/staff call
- Routine call (sub station call-ins on standby connected according to urgency level and time of call)
- Warning Tones
- Line-off trouble at bedside or NHR-3TS and duty station - to stop tone on duty station, plug in 2-pin jumper to CN104 in Central Control Unit
- 90 sec. elapse timer - call tone goes up to HIGH level
- 90 sec. off-hook timer - steady tone will notify handset is off-hook
- Wire Type:
- Master Station to Central Control Unit: 5 twisted pair, low cap, PE, overall shield - use Aiphone #862210
- Distance: Master Station to Central Control Unit: 65'(22AWG)
- Dimensions: 8-1/4 in H x 11 in W x 3-3/4 in D